Ben Lewis, Zoey Z. Hu, Dahyeon Lee, Kyungtae Kim, William Warfield, Kai Zhou, Daniele Nicolodi, Thomas Legero
Sets a new record for laser stability (7×10⁻¹⁸) with convincing thermal-noise-limited evidence and direct clock validation, giving a benchmark and enabling tool for the optical clock community, though the concept is incremental.
We report a laser with record frequency stability, based on a 21 cm cryogenic silicon cavity. For best short-term stability, the cavity operates at 6 K to minimize Brownian thermal noise and achieves a fractional stability of between 2 to 50 s of averaging time and 2.5 mHz linewidth. For best long-term stability the cavity operates at 16 K, where silicon has a zero-crossing of thermal expansion. Slightly increased thermal noise limits the performance to , consistent with the expected temperature scaling, but the long-term performance is improved. The laser then has a linewidth of 2.9 mHz and long-term frequency drift less than 4 μHz s. The laser is used to enhance the performance of a state-of-the-art Sr lattice clock, operating with Rabi and Ramsey interrogation times beyond 10 s.
Paper type: Experimental/tool-system paper in precision metrology (ultrastable lasers).
The paper reports a cryogenic silicon cavity laser (Si7, 21 cm) with a fractional frequency instability of 7×10⁻¹⁸ at 6 K and 9×10⁻¹⁸ at 16 K, a new record, flat between about 2 and 50 s. Linewidths are 2.5 and 2.9 mHz, and long-term drift is below 4 µHz/s. The advance comes from two design changes to the earlier JILA/PTB 21 cm silicon cavity:
Together these cut thermal noise three- to fivefold. Enabling engineering includes:
The laser is demonstrated on a Sr lattice clock with Ramsey dark times up to 12 s and a 12.4 s Rabi pulse.
The stability measurement uses a three-cornered hat against two other silicon cavities (Si3, Si6) with Ekstrom–Koppang error bars. An independent check comes from the Sr clock for long-term behaviour. This is the standard approach, and the cross-validation is good.
The thermal-noise attribution is persuasive:
Weaknesses:
Direct industrial translation is limited. The system is a complex, expensive, bespoke laboratory apparatus, and the audience is the national-metrology-lab and frontier-AMO community.
The field is in the "laser stability is the bottleneck" regime. Clock systematics have reached 10⁻¹⁸–10⁻¹⁹, and competing efforts include room-temperature long cavities at 4×10⁻¹⁷, a 4 K 10 cm cavity at 4×10⁻¹⁷, and crystalline-coated silicon at 2.5×10⁻¹⁷. A result well below 10⁻¹⁷ is timely, and it was likely the benchmark the community awaited. Redefinition of the SI second also raises the demand for such oscillators.
This is a high-quality, record-setting metrology paper from the leading groups (JILA, PTB). It is not conceptually revolutionary, but it sets the new benchmark in ultrastable lasers and materially enables improved optical clocks and fundamental-physics searches. It is likely to be widely cited as the state-of-the-art reference. Its impact is concentrated in the precision-measurement community, with real downstream value in clock networks, timekeeping and tests of fundamental physics.
Generated Oct 23, 2026
Sets a new record for laser stability (7×10⁻¹⁸) with convincing thermal-noise-limited evidence and direct clock validation, giving a benchmark and enabling tool for the optical clock community, though the concept is incremental.